ZHANG Xing-yuan, LONG Wei, LU Bin, et al. Design of a high sensitivity liquid level detection system and its clinical application[J]. 2014, 33(6): 72-75+79.
ZHANG Xing-yuan, LONG Wei, LU Bin, et al. Design of a high sensitivity liquid level detection system and its clinical application[J]. 2014, 33(6): 72-75+79. DOI: 10.13873/j.1000-97872014.06.008.
Design of a high sensitivity liquid level detection( LLD) system and its application in automatic biochemical analyzer is introduced. Contact capacitive method principles is regarded as core
a strong insulated metal double casing sampling needle is used as detection sensor
and a PLL phase comparator-based capacitance detection circuit combined with adaptive recognition algorithm of effective pulses is applied in the system
as a result
a performance indicators like cross-contamination rate is not more than 0. 3 % and as minimum sample is 2 μL
sampling error is not more than ± 3 % are achieved by the system. Automatic biochemical analyzer clinical application shows that the system has characteristics such as high detection sensitivity