LI Xiao-yan, YANG Hong, ZENG Xiang-hua, et al. Study of SPR sensing detection method for CRP in blood[J]. 2013, 32(12): 38-41. DOI: 10.13873/j.1000-97872013.12.028.
A novel method for rapidly and accurately analyzing C reactive proteins( CRP) in blood samples is established by using surface plasmon resonance( SPR) biosensing technology. Continuous angle-scanning mode is adopted in the Kretchmann-configuration SPR biosensor
and two individual step motors are used to drive rotation of a prism and a photoelectric detector for high-precision analysis of a single sample. Anti-CRP antibody molecules are coated on the surface of the sensor chip
detect and analyze concentration of CRP in blood samples with SPR sensor. SPR and immune turbidimetry methods are used to detect standard CRP sample and 200 blood samples from children( 7
10 years old
112 men and 88 women) with infectious diseases
and results show that SPR method detects standard sample has wider linear range. In detection of blood samples from children
SPR biosensor can provide a similar result as that from immune turbidimetry method. However
SPR method is quicker
lower consumption
and better repeatability. Thus
SPR biosensing analysis method has outstanding advantages in CRP detection
and may be widely used in clinical detection analysis.