A chip for blood coagulation detection is designed based on the principle of magnetoelastic transducer.The chip is composed of Fe-based amorphous alloys and an inverted T-shaped plastic bottom film. The dynamic tracking of the blood clotting process and the determination of prothrombin time are achieved by using the influence of liquid viscosity changing on resonance characteristics of Fe-based amorphous alloy. Experimental results show that amplitude response of the detection chip has high degree of correlation with blood coagulation process. The mean value of prothrombin time in three tests is 13. 7 s determined by the maximum point of the slope of the amplitude
the measurement result of the artificial tube method is 13. 1 s
which indicates that this method has high accuracy. In addition
this kind of chip is easy to fabricate
low cost
small in size
and a very small quantity of blood is required for detected.
关键词
Keywords
references
嵌入式血液凝固检测系统的设计 [D]. 赵正卿. 武汉理工大学 2012
Comparison of the chemical profiles and anti-platelet aggregation effects of two “Dragon’s Blood” drugs used in traditional Chinese medicine [J] . Tao Yi,Hu-Biao Chen,Zhong-Zhen Zhao,Zhi-Ling Yu,Zhi-Hong Jiang. Journal of Ethnopharmacology . 2010 (2)
Surface plasmon resonance detection of blood coagulation and platelet adhesion under venous and arterial shear conditions [J] . Biosensors and Bioelectronics . 2007 (2)
Monitoring complex formation in the blood-coagulation cascade using aptamer-coated SAW sensors [J] . T.M.A. Gronewold,S. Glass,E. Quandt,M. Famulok. Biosensors and Bioelectronics . 2004 (10)