CUI Yuan-yuan, WANG Bo-xiong, LIU Jian-nan, et al. Time-of-flight measurement system based on ultrasonic wave gas transducer[J]. 2013, 32(6): 98-100+107.
CUI Yuan-yuan, WANG Bo-xiong, LIU Jian-nan, et al. Time-of-flight measurement system based on ultrasonic wave gas transducer[J]. 2013, 32(6): 98-100+107. DOI: 10.13873/j.1000-97872013.06.038.
A time-of-flight measurement system based on bi-echo method and digital correlation method is proposed and built.Bi-echo method is used to calibrate circuit delay of sensor and obtain reference signal for measurement.Take the maximum location of two-layer correlation function between reference signal and receiving signal as coarse location point
by linear interpolating the Hilbert transform function around this coarse location point
the zero-crossing point is acquired to be the final measurement value of time-of-flight.The calibration precision of the system can reach fs level
and the single measurement precision can reach a few ns.The system can obtain reliable and precise ultrasonic time-of-flight
and can be applied to the fields of distance
flow and wind velocity measurement.
关键词
Keywords
references
FPGA-based High-precision Measurement Algorithm for the Ultrasonic Echo Time of Flight [J]. 王伯雄,张金. Journal of Measurement Science and Instrumentation . 2010(02)
High-frequency transducers and correlation method to enhance ultrasonic gas flow metering [J] . P Brassier,B Hosten,F Vulovic. Flow Measurement and Instrumentation . 2001 (3)