ZHANG Xiao-juan, ZENG Xiang-hua, LIU Lin-lin, et al. Simulation of flow field in micro flow pool in SPR sensor array[J]. 2013, 32(4): 83-87. DOI: 10.13873/j.1000-97872013.04.037.
In research and fabrication of micro flow pool for a self-made SPR sensor array device
in order to achieve high precision and repeatability
flow through each detecting point is required as coincident as possible.Several different flow pool structures are designed and simulated by using Comsol Multiphysics software.At first
inner flow field distribution of common shuttle-shape flow pool is analyzed
and impact of different transition section is analyzed.Based on this simulation
impact of different shapes of flow pools
especially their outlines
on the flow field distribution are also discussed.And then
effect of some microstructures array design on flow field distribution of detecting area.Simulation results show that appropriately extending the length of the transition section will help to improve the stability and uniformity of the flow of fluid.However
the improvement become smaller when the length exceeded a certain value.While effect of different shapes
on flow field of detecting area is not obvious.Microstructure array design can bring some improvement on the homogeneity of the flow field.
Development of an ultra‐low volume flow cell for surface plasmon resonance detection in a miniaturized capillary electrophoresis system [J] . Attila Gaspar,Frank A. Gomez. ELECTROPHORESIS . 2012 (12)