LIANG Chun-yan1, TIAN Xue-long1. Design of pressure algometer based on pull or press force sensor[J]. 2013, 32(3): 104-106+109. DOI: 10.13873/j.1000-97872013.03.040.
a kind of algometer based on high precision pull or press force sensor is developed.This instrument uses single-chip microcomputer C8051F020 as core
and by signal acquisition and processing circuit coordinated with software programming
collection
display and data storage of pressure pain threshold(PPT)are realized.Test results show that the obtained PPT has higher accuracy(<0.3N)and better stability
which provides reliable technical support for treatment and curative effect evaluation.
Test-Retest Reliability of Pressure Pain Threshold Measurements of the Upper Limb and Torso in Young Healthy Women [J] . David H. Jones,Robert D. Kilgour,Alain S. Comtois. Journal of Pain . 2007 (8)