CHEN Ji, FANG Dian-dian, ZHAO Xiao-ming, et al. Research and fabrication of noncontact tonometer based on principle of force equilibrium[J]. 2013, 32(1): 91-94+101.
CHEN Ji, FANG Dian-dian, ZHAO Xiao-ming, et al. Research and fabrication of noncontact tonometer based on principle of force equilibrium[J]. 2013, 32(1): 91-94+101. DOI: 10.13873/j.1000-97872013.01.010.
based on force balance principle and air static pressure technology.Overall structure of a non-contact tonometer design of intraocular pressure measurement module and frame type air cushion are introduced.The intraocular pressure measurement module includes nozzle
corneal alignment system and a corneal applanation photoelectric detecting system.Frame type air cushion adopts the orifice of aerostatic thrust bearing.The pressure distribution of external frame type air cushion
and test the nozzle and corneal photoelectric detecting system are calculated and simulated.The experimental results show that when distance between cornea and nozzle is rated distance L
nozzle gas can be flat corneas
photoelectric receiving pipe just receives maximum light.Compared with imported instruments
absolute error of measuring mean value is 1.3 mmHg
development of noncontact tonometer can realize noncontact measurement.
New ways to measure intraocular pressure [J] . Mohammed K ElMallah,Sanjay G Asrani. Current Opinion in Ophthalmology . 2008 (2)
Dynamic Contour Tonometry: Presentation of a New Tonometer [J] . Hartmut E Kanngiesser,Christoph Kniestedt,Yves C A Robert. Journal of Glaucoma . 2005 (5)