TENG Peng-fei1, CHENG Jia2, YANG Qi-hua1, et al. Study on piston calibrating device for infusion pump analyzer[J]. 2012, 31(12): 16-18+22. DOI: 10.13873/j.1000-97872012.12.003.
活塞式输液泵分析仪检定装置的研究
摘要
为了解决国内很少有流量标准装置可以检定输液泵分析仪的问题
研制了一种基于主动活塞式的输液泵分析仪检定装置。阐述了检定装置的工作原理、结构
设计了集成数据采集和自动控制的系统。检定装置利用上位机控制工作流程
采集瞬时流量等信号。对输液泵分析仪在0.1~10 mL/h流量范围内进行了检定实验
证明分析仪的精度达到2%。同时分析了影响输液泵分析仪流量检定的不确定度因素
其扩展不确定为2.4%(k=2)。结果表明:该装置具有良好的实用性和广阔的应用前景。
Abstract
In order to solve problem that few flow standard device is able to test infusion pump analyzer at present
a apparatus active piston based is proposed.The principle and structure of the calibrating device is described.A system for data acquisition and automatic control is designed.An upper computer is used to control workflow and acquire instantaneous flow and other signals.The verification experiment proves that
for flow range 0.1
1
0 mL/h
the precision of the analyzer reaches 2 %.Uncertainty factors that affect flow calibration of infusion pump analyzer are analyzed and the extension uncertainty equal to 2.4 %(k=2).The result shows that this apparatus has characters of good practicability and wide application prospect.