GAO Hong-tao1, DU Juan1, XIAO Nan2, et al. Research on acquisition and control of information for force sensors in catheter operating system[J]. 2012, 31(1): 82-84+89.
GAO Hong-tao1, DU Juan1, XIAO Nan2, et al. Research on acquisition and control of information for force sensors in catheter operating system[J]. 2012, 31(1): 82-84+89. DOI: 10.13873/j.1000-97872012.01.032.
acquisition and control for force feedback of force sensors have been studied
the top force of catheter is measured using fiber-optic sensor.The force of catheter wall is measured by micro force sensor based on PVDF material.The catheter operating system monitors output information of force sensors
and a novel lord-from catheter operating system with force information monitoring is achieved.Experimental results show that the system can realize the feedback of the force information of sensors very well.
关键词
Keywords
references
A biomimetic sensor for a crawling minirobot [J] . Weiting Liu,A. Menciassi,S. Scapellato,P. Dario,Yuquan Chen. Robotics and Autonomous Systems . 2006 (7)
Design and fabrication of a largely deformable sensorized polymer actuator [J] . Jaewook Ryu,Jungyul Park,Byungkyu Kim,Jong-Oh Park. Biosensors and Bioelectronics . 2005 (5)