将葡萄糖氧化酶(GOD)固定在多壁碳纳米管(MWCNTs)修饰电极(ME)上
GOD催化氧化葡萄糖生成过氧化氢
并使鲁米诺产生电致化学发光(ECL)
据此构建了一种新型ECL葡萄糖传感器。结果表明:CNTs修饰的电极对鲁米诺和H2O2反应具有显著的电催化活性和增敏效果。该传感器对葡萄糖检测的线性范围为0.01~10.0mmol/L
相关系数R=0.999(n=5)
检测限为5.0μmol/L。此传感器响应快
稳定性高
测定条件接近人体自然pH。将所建立的方法用于临床血清样品中葡萄糖含量的测定
获得了满意的结果。
Incorporation of multi-walled carbon nanotubes on the electrode is performed and modification of glassy carbon electrode with the prepared nano-hybrid material led to the fabrication of a novel electrode.The modified electrode shows attractive electrocatalytic activity and sensitizing effect on Luminol-H2O2 electrochemiluminescence(ECL) reactions at neutral media.The sensitized Luminol-H2O2 reactions are successfully applied for the ECL determination of glucose.Under the optimal conditions for Luminol-H2O2 system
the ECL signal intensity of luminol is linear with the concentration of glucose in the range of 0.01~10.0 mmol/L
correlation coefficient R=0.999(n=5)
the limits of detection for glucose are 5.0 μmol/L.Excellent electrocatalytic activity
high stability
technically simple and possibility of preparation at short period of time are of great advantages of this glucose biosensor.
Electrogenerated chemiluminescence biosensor for glucose based on poly(luminol–aniline) nanowires composite modified electrode [J] . Guixin Li,Jinlan Lian,Xingwang Zheng,Jing Cao. Biosensors and Bioelectronics . 2010 (2)
Adsorption of glucose oxidase at platinum-multiwalled carbon nanotube-alumina-coated silica nanocomposite for amperometric glucose biosensor [J] . Ming-Chieh Tsai,Yu-Chen Tsai. Sensors & Actuators: B. Chemical . 2009 (2)
One-step synthesis of silver nanoparticles/carbon nanotubes/chitosan film and its application in glucose biosensor [J] . Jiehua Lin,Chunyan He,Yue Zhao,Shusheng Zhang. Sensors & Actuators: B. Chemical . 2009 (2)
Amperometric glucose biosensor with high sensitivity based on self-assembled Prussian Blue modified electrode [J] . Yu Liu,Zhenyu Chu,Yannan Zhang,Wanqin Jin. Electrochimica Acta . 2009 (28)
A highly performing electrochemiluminescent biosensor for glucose based on a polyelectrolyte-chitosan modified electrode [J] . Hong Dai,Xiaoping Wu,Huifeng Xu,Youmei Wang,Yuwu Chi,Guonan Chen. Electrochimica Acta . 2009 (19)
Glucose biosensor based on gold nanoparticle-catalyzed luminol electrochemiluminescence on a three-dimensional sol–gel network [J] . Xiaoqing Liu,Wenxin Niu,Haijuan Li,Shuang Han,Lianzhe Hu,Guobao Xu. Electrochemistry Communications . 2008 (9)
Hollow spherical nanostructured polydiphenylamine for direct electrochemistry and glucose biosensor [J] . P. Santhosh,K.M. Manesh,S. Uthayakumar,A.I. Gopalan,K.-P. Lee. Biosensors and Bioelectronics . 2008 (7)
Clinical determination of glucose in human serum by a tomato skin biosensor [J] . Hui Han,Yi Li,Huan Yue,Zaide Zhou,Dan Xiao,Martin M.F. Choi. Clinica Chimica Acta . 2008 (1)
A novel glucose biosensor based on immobilization of glucose oxidase in chitosan on a glassy carbon electrode modified with gold–platinum alloy nanoparticles/multiwall carbon nanotubes [J] . Xinhuang Kang,Zhibin Mai,Xiaoyong Zou,Peixiang Cai,Jinyuan Mo. Analytical Biochemistry . 2007 (1)
Glucose biosensor prepared by glucose oxidase encapsulated sol-gel and carbon-nanotube-modified basal plane pyrolytic graphite electrode [J] . Abdollah Salimi,Richard G. Compton,Rahman Hallaj. Analytical Biochemistry . 2004 (1)
A novel multi-walled carbon nanotube-based biosensor for glucose detection [J] . S.G Wang,Qing Zhang,Ruili Wang,S.F Yoon. Biochemical and Biophysical Research Communications . 2003 (3)
A novel flow through optical fiber biosensor for glucose based on luminol electrochemiluminescence [J] . Liande Zhu,Yingxiu Li,Guoyi Zhu. Sensors & Actuators: B. Chemical . 2002 (2)
Electrochemiluminescent determination of glucose with a sol–gel derived ceramic–carbon composite electrode as a renewable optical fiber biosensor [J] . Liande Zhu,Yingxiu Li,Faming Tian,Bo Xu,Guoyi Zhu. Sensors & Actuators: B. Chemical . 2002 (2)