WANG Hai-tao, LIN Yu-chi, FU Lu-hua, et al. Experiment research on measuring system for micro-force value[J]. 2009, 28(11): 8-11. DOI: 10.13873/j.1000-97872009.11.026.
A design method of measuring system for micro-force value is introduced.Self-designed sensor structure is used to make full use of a cylindrical capacitance sensor range
linear
the edge of the characteristics of small effect and so on
a straight line is used through the high-precision bearings to ensure the movement of coaxial
and other methods of shield capacitance are used to eliminate the interference.Detection circuit through the new high-precision
integrated capacitance-to-digital converter and the AD7747 single-chip processor are used to achieve accurate measurement of dC changes in the volume while reducing the cost of the system.The experiments conclude that the method for the following 10-5N power source and measuring the value provides an effective solution.