LI Jin, ZHENG Xiao-lin, HOU Wen-sheng, et al. Error correction of three dimensional magnetic localization system used for tracking micro diagnosis capsule in gastrointestinal tract[J]. 2008, 27(11): 93-96+99.
LI Jin, ZHENG Xiao-lin, HOU Wen-sheng, et al. Error correction of three dimensional magnetic localization system used for tracking micro diagnosis capsule in gastrointestinal tract[J]. 2008, 27(11): 93-96+99. DOI: 10.13873/j.1000-97872008.11.034.
A 3-axis magnetic inductive sensor array system used for localization and tracking of micro diagnosis capsule in human gastrointestinal tract is designed
error sources of this tacking system are analyzed
and the error correction methods are proposed to obtain better accuracy.The correction methods include background magnetic field elimination
sensitivity calibration
sensor position correction and sensor orthonormality adjustment.The experiment results show that the error correction procedure can improve the localization accuracy prominently.For the 8mm×8mm cylindrical N35 NdFeB magnet
the average position error is less than 8mm
and orientation error is less than 6°when the magnet is 150mm above the sensor array plane after the error correction procedure is implemented.The localization system can also track the locus of the capsule with a magnet inside
when it moves among the simulative intestinal tract.