XIONG Xing-liang1, CAI Shao-xi2, XIE Guo-ming1. Study on endotoxin piezo-biosensor based on nano-Au self-assembled monolayer[J]. 2007, (6): 29-32. DOI: 10.13873/j.1000-97872007.06.026.
An endotoxin piezo-biosensor is developed by using nano-Au self-assembled monolayer(SAM) techniques.On a surface of quartz crystal microbalance(QCM)
an ethanthiol(HS-(CH2)6-SH) SAM is formulated to assemble a nano-Au monolayer by covalent bond.Thus
the modified interface used to immobilize polymyxin B(PMB) to detect endotoxin(ET) in water.The main conditions of assembly of nano-Au and immobilization of PMB are investigated.Compared with SAM without nano-Au immobilization
the sensitivity and discrimination of the proposed biosensor are improved.
Enhancement in the sensitivity of a gas biosensor by using an advanced immobilization of a recombinant bioluminescent bacterium [J] . Geun Cheol Gil,Young Joon Kim,Man Bock Gu. Biosensors and Bioelectronics . 2002 (5)
Piezoelectric quartz crystal biosensors [J] . Renee L. Bunde,Eric J. Jarvi,Jeffrey J. Rosentreter. Talanta . 1998 (6)
Capacitive monitoring of protein immobilization and antigen–antibody reactions on monomolecular alkylthiol films on gold electrodes [J] . Vladimir M Mirsky,Michael Riepl,Otto S Wolfbeis. Biosensors and Bioelectronics . 1997 (9)