LIU Lin-wei, DENG Wen-di. Implementation of AC millivoltmeter based on equivalent sampling[J]. 2018, (10): 92-94. DOI: 10.13873/j.1000-9787(2018)10-0092-03.
基于等效采样的交流毫伏表实现
摘要
通过对等效随机采样和等效顺序采样2种等效采样方式的分析对比
提出了一种10 MHz交流毫伏表的设计方法。前端调理电路将待测信号衰减或放大到合适的幅度
现场可编程门阵列(FPGA)驱动模/数转换器(ADC)芯片实现等效采样
单片机(MCU)对采样所得数据进行分析处理
得到最终测量值。经过测试
采用1 MHz采样率的ADC芯片
最终可以还原10 MHz以下的待测信号
并能够精准测得其有效值。
Abstract
Through analysis and comparison of equivalent random sampling and equivalent sequential sampling
present a design proposal for 10 MHz AC millivoltmeter. The fount-end adjusting circuit modulates the signal to an appropriate amplitude
then field programmable gate array( FPGA) drives the ADC chip to achieve equivalent sampling
the final measured value is calculated after microprogrammed control unit( MCU) processing the sampled data. Test results shows that the signal under 10 MHz can be restored and the effective value of the signal can be measured precisely by using the ADC chip of 1 MHz sampling frequency.