上海交通大学微纳科学技术研究院微米/纳米加工技术国家重点实验室薄膜与微细技术教育部重点实验室
纸质出版:2024
移动端阅览
王卓, 刘景全. 集成PVB修饰Ag/AgCl参比电极的pH传感硅探针阵列[J]. 传感器与微系统, 2024,(9):68-70.
王卓, 刘景全. 集成PVB修饰Ag/AgCl参比电极的pH传感硅探针阵列[J]. 传感器与微系统, 2024,(9):68-70. DOI: 10.13873/J.1000-9787(2024)09-0068-03.
介绍了一种带有集成Ag/AgCl参比电极的硅记录探针,通过表面聚乙烯醇缩丁醛(PVB)的修饰实现稳定的pH传感,用于应用神经科学。集成参比电极通过恒流电化学镀法实现银层的制备以及氯化,滴涂的PVB主要用作参比电极的保护膜,参比电极面积为0.44 mm
2
。最终硅记录探针的pH灵敏度为53.13 mV/pH
电位漂移为8 000 s下20 mV
并且探针工艺与硅基探针高度融合。总之,集成参比电极并且提高其稳定性对于生物传感器以及脑疾病诊断生物标志物的电化学检测具有重要意义。
Department of Electronic and Computer Engineering, The Hong Kong University of Science and Technology,kcetjon@connect.ust.hk.,Department of Electronic and Computer Engineering, The Hong Kong University of Science and Technology..Fabrication of Thin Film Silver/Silver Chloride Electrodes with Finely Controlled Single Layer Silver Chloride.[J].Journal of visualized experiments : JoVE,2020.
Kai Chun Eddie Tjon,Jie Yuan.Impedance characterization of silver/silver chloride micro-electrodes for bio-sensing applications[J].Electrochimica Acta,2019.
Bufeng Bao,Bowen Ji,Minghao Wang,Kunpeng Gao,Bin Yang,Xiang Chen,Xiaolin Wang,Jingquan Liu.Development and characterisation of electroplating silver/silver chloride modified microelectrode arrays[J].Micro & Nano Letters,2019.
Nimbalkar Surabhi,Castagnola Elisa,Balasubramani Arvind,Scarpellini Alice,Samejima Soshi,Khorasani Abed,Boissenin Adrien,Thongpang Sanitta,Moritz Chet,Kassegne Sam.Ultra-Capacitive Carbon Neural Probe Allows Simultaneous Long-Term Electrical Stimulations and High-Resolution Neurotransmitter Detection.[J].Scientific reports,2018.
Munawar Khalil,Sai Wang,Jianjia Yu,Robert L. Lee,Ning Liu.Electrodeposition of Iridium Oxide Nanoparticles for pH Sensing Electrodes[J].Journal of The Electrochemical Society,2016.
M. Tabata,C. Ratanaporncharoen,A. Asano,Y. Kitasako,M. Ikeda,T. Goda,A. Matsumoto,J. Tagami,Y. Miyahara.Miniaturized Ir/IrOx pH Sensor for Quantitative Diagnosis of Dental Caries[J].Procedia Engineering,2016.
Vreeland Richard F,Atcherley Christopher W,Russell Wilfred S,Xie Jennifer Y,Lu Dong,Laude Nicholas D,Porreca Frank,Heien Michael L.Biocompatible PEDOT:Nafion composite electrode coatings for selective detection of neurotransmitters in vivo.[J].Analytical chemistry,2015.
Mehdi Jorfi,John L Skousen,Christoph Weder,Jeffrey R Capadona.Progress towards biocompatible intracortical microelectrodes for neural interfacing applications[J].Journal of Neural Engineering,2015.
T Matsumoto,A Ohashi,N Ito.Development of a micro-planar Ag/AgCl quasi-reference electrode with long-term stability for an amperometric glucose sensor[J].Analytica Chimica Acta,2002.