ZHOU Zhi, SU Xiaoguang, LI Ming, et al. Funnel-type dual-electrode chloride ion concentration detection method in liquid/gas phase conditions[J]. 2021, (12): 149-152.
ZHOU Zhi, SU Xiaoguang, LI Ming, et al. Funnel-type dual-electrode chloride ion concentration detection method in liquid/gas phase conditions[J]. 2021, (12): 149-152. DOI: 10.13873/J.1000-9787(2021)12-0149-04.
液相/气相条件漏斗型双电极氯离子浓度检测方法
摘要
针对液相/气相环境氯离子浓度检测需求
提出了一种基于双电极检测体系的漏斗状氯离子浓度检测方法。通过设计漏斗型收集单元
实现对气相环境样品有效获取。根据传感器输出电势差
计算收集单元沉积液中氯离子浓度。借助3D打印技术制作聚四氟乙烯外壳
采用电镀法制得Ag/AgCl电极。在1×10
-3
~1 mol/L氯化钠浓度范围的液相条件下
校准曲线线性度约为99.14%
灵敏度达到54.63 mV/lgC
Cl
-
检测下限为0.891×10
-4
mol/L
呈现较好的灵敏度和检测区间。在沉降速率为100 mL/(80 cm
2
·h)的气相环境下
基于漏斗型收集单元的传感器峰值响应时间约为440 s
而未采用漏斗型收集单元设计则无有效数据。
Abstract
Aiming at the requirement of chloride ion concentration detection in liquid/gas phase environment
a funnel-shaped chloride ion concentration detection method based on two-electrode detection system is proposed.By designing a funnel-type collection unit
effective acquisition of gas phase environmental samples is achieved.According to the output potential difference of the sensor
the chloride ion concentration in the sediment of the collection unit is calculated.A 3 D printing technique is used to make a polytetrafluoroethylene shell
and an Ag/AgCl electrode is prepared by electroplating.Under th
e liquid phase condition of 1×10
-3
~1 mol/L sodium chloride concentration range
the linearity of the calibration curve is about 99.14 %
the sensitivity reaches 54.63 mV/lgC
Cl
-
and the lower detection limit is 0.891×10
-4
mol/L.Good sensitivity and detection interval.In a gas phase environment with a settling rate of 100 mL/(80 cm~2 · h)
the peak response time of the sensor based on the funnel-type collection unit is about 440 s