Lü Shuyuan, WANG Feng, LIU Chongqi. Research on CO2 detection system based on TDLAS-WMS[J]. 2021, (3): 52-54+57. DOI: 10.13873/J.1000-9787(2021)03-0052-03.
基于TDLAS-WMS的CO2检测系统研究
摘要
采用空心光子晶体光纤(HC-PCF)作为气室
将可调谐二极管激光光谱吸收技术—波长调制技术(TDLAS-WMS)与谐波比值检测相结合
利用CO
2
吸收光谱中1 572.335 nm吸收峰
实现了CO
2
气体体积分数测量。系统采用发射波长为1573 nm
功率为1 m W的分布反馈式激光器(DFB)作为光源
5 k Hz余弦波用于波长调制
20 Hz锯齿波用于波长扫描
使用光电探测器、数据采集卡采集被气体吸收的光信号
采用基于Lab VIEW的后处理相关算法计算CO
2
气体体积分数。实验结果表明:在1 s内测量系统最小探测极限为25×10
-6
在31 s内可进一步提高至15×10
-6
。
Abstract
The volume fraction of CO
2
gas is detected using hollow-core photonic crystal fiber( HC-PCF) as gas cell
tunable diode laser absorption spectroscopy and wave length modulation spectroscopy( TDLAS-WMS) and harmonic ratio detection method are combined and use absorption peak of 1 572. 335 nm in CO
2
absorption spectrum. A 1 m W distribute feedback( DFB) laser with an emission wavelength of 1 573 nm is used as the light source in this system. 5 k Hz cosine signal is used to wavelength modulation and 20 Hz sawtooth wave wavelength scanning. The absorbed light is detected with a photo detector and a data acquisition card. The post-proces
sing correlation algorithm based on Lab VIEW is used to calculate the CO
2
gas volume fraction. Experimental results indicate that the minimum detection limit of the detection system is 25 × 10