ZHANG Yunpeng, WANG Jiping, XIONG Daxi, et al. Design of wearable wireless bioelectric feedback control system[J]. 2021, (2): 73-75+82. DOI: 10.13873/J.1000-9787(2021)02-0073-03.
可穿戴无线生物电反馈控制系统设计
摘要
针对基础医学中对脑电、肌电、心电等多种生物电采集及反馈的需要
采用高集成度的模拟前端作为信号传感方案
提出了一种基于生物电的闭环反馈控制方法
并给出具体电路实现和系统设计。实验结果表明:本系统信号采集范围在±2.5 V之间
精度可以达到0.5μV
与标准仪器对比结果一致;生物电反馈范围±5 V之间
精度可以达到0.1 V;闭环控制实验表明本系统可实时反馈可调电刺激。本系统尺寸为14 cm×6 cm×2 cm
重量292 g
结合5000 m Ah锂电池
系统可连续工作一周时间
其无线空旷控制距离45 m
满足多类型生物电采集、无线可调电脉冲反馈
以及闭环实时控制的需要。
Abstract
Aiming at the needs of a variety of bioelectricity acquisition such as electroencephalogram( EEG)electrocardio signals and feedback in basic medicine
use highly integrated analog front-end scheme
a closed-loop feedback control method based on bioelectricity is proposed
and specific circuit implementation and system design is given. The experimental results show that the signal acquisition range of this system is between ± 2. 5 V
and the precision can reach 0. 5 μV
which is consistent with the results of standard instruments. The precision of bioelectric feedback is 0. 1 V in range of ± 5 V. Closed-loop control experiments show that the system can feedback adjustable electrical stimulation in real time. The system has a size of 14 cm × 6 cm × 2 cm and a weight of 292 g.Combined with 5 000 m Ah lithium battery
the system can work continuously for one week. Its wireless spacious control distance is 45 m
which meets the needs of multi-type bioelectricity acquisition
wireless adjustable electric pulse feedback and closed-loop real-time control.