LIU Zhaoxu, WANG Minghao, GUO Zhejun, et al. Miniaturized design of implantable brain-computer interface system[J]. 2019, 38(9): 74-76+80. DOI: 10.13873/J.1000-9787(2019)09-0074-03.
low-cost and portable brain-computer interface system for implantable neural signals acquisition is developed. This system mainly includes front-end neural signal acquisition module
signal conversion module
controller module
data transmission module and power management module. Besides
host computer is used to implement real-time data flow receiving and visual graphical interface. Experimental results show that the 16-channel brain-computer interface system can achieve sampling frequency of 20 kHz per channel
sampling precision of 16 bits and system input noise of 7 μV
closing to the performance of the mainstream acquisition equipment. The system meets the requirements of multi-channel
high sampling rate
high precision
and low noise neural signal acquisition for implantable brain-computer interface.