YAN Ning, ZHAO Yanzheng, YAN Weixin. Analysis and optimization of separation liquid path control system for circulating tumor cell[J]. 2019, (5): 67-70.
YAN Ning, ZHAO Yanzheng, YAN Weixin. Analysis and optimization of separation liquid path control system for circulating tumor cell[J]. 2019, (5): 67-70. DOI: 10.13873/J.1000-9787(2019)05-0067-04.
In response to the growing demand for in-vitro diagnostics
the need for personalized care and the need for early detection of cancer
use difference circulating tumor cell( CTC) and normal cell physical characteristics
through the microporous membrane and corresponding separation and enrichment of liquid path
the physical separation of specific cells and normal cells is realized. The liquid path pressure is simulated and analyzed by Fluent software. Aiming at the case that in the process of the liquid flow
the microporous membrane for blood component plug hole induced porosity decreases
causing the pressure into the microporous membrane bear a reasonable interval
and even lead to rupture of membranes blocked. A one-dimensional fuzzy control rule is proposed
which can make the pressure of the microporous filter stable within a reasonable range.