GAI Sha-sha, MAO Wen-hua, ZHANG Jun-ning, et al. Design of signal source for NIR spectrometer based on AOTF[J]. 2016, 35(12): 103-105+109. DOI: 10.13873/J.1000-9787(2016)12-0103-03.
Acousto-optic tunable filter( AOTF) is the fifth generation of spectroscopic system core component. In AOTF near-infrared( NIR) spectrometer system. Stable RF driven signal source is a key issue in research and development of AOTF NIR Spectrometer system. Based on wavelength range of near infrared spectroscopy
The system uses SCM to drive DDS and designs output frequency range at 80
150 MHz. The system includes a singlechip control module
DDS signal generator module
filter circuit
power amplifier module. Test results show that in the desired output frequency range
the system can output the desired waveform and output signal frequency relative error within 0. 45 %
with power of 1W
corresponding to the output wavelength range of 900