MA Min, LU Cheng-chao, JIANG Feng. Analysis on signal integrity based on ECT system[J]. 2016, 35(7): 30-31+34. DOI: 10.13873/J.1000-9787(2016)07-0030-02.
In order to further improve measuring precision of electrical capacitance tomography( ECT) acquiring system
analyze and research on effect of integrity of ECT signal on measurement results. Aiming at characteristics of weak measuring capacity signal
and the value is in p F level
an ECT acquisition system designed considering signal integrity
duty ratio of sensor is 81 %. Then
the above system is used to analyze the effect of signal integrity on measured results. Experimental results show that when the peak-to-peak value of sine excitation signal is18. 7 V
after current voltage( C/V) conversion
the voltage peak-to-peak value of the collected biggest signal is921 m V
which is consistent with the theoretical calculation results; by using sensitivity matrix coefficient algorithm
clear ECT images are obtained on PC. The results show that the system meets the requirements of measurement precision of system.