LIU Ya, LUO Zhi-hui, XU Xiao-ming, et al. Three axis weak magnetic field detection system based on Lab VIEW[J]. 2015, 34(7): 73-75+78. DOI: 10.13873/J.1000-9787(2015)07-0073-03.
According to requirement of weak magnetic field detection
Lab VIEW virtual instrument technology developed three axis of weak magnetic field detection system
using 3D Mag IC( PNI12927) chip to drive the three axis magnetic sensor
configuration parameters and communication control on sensing device are carried out by STC12LE2052 AD MCU
and multi-function visualization system software is compiled based on Lab VIEW. By calibrating the sensitivity and linearity of the detection system
the systematic error of which is pre-compensated by means of software
real-time measurement of magnetic field at the level of 15 n T is realized. Ceomagnetic field measurement results verify the feasibility of the method.
A method for real-time compensation of moving ferromagnet’s magnetic moment [J] . Zhou Jian-jun,Lin Chun-sheng,Fu Kang. Journal of Magnetism and Magnetic Materials . 2013