LI Li, WU Jiahui, ZHAI Liying, et al. Preparation of H2V12O31/BiOI/FTO photoelectrochemicalsensor and its detection of tetracycline hydrochloride[J]. 2026, 45(1): 172-178.
LI Li, WU Jiahui, ZHAI Liying, et al. Preparation of H2V12O31/BiOI/FTO photoelectrochemicalsensor and its detection of tetracycline hydrochloride[J]. 2026, 45(1): 172-178.DOI:
)is prepared by sol-gel method and deposited onto the surface of bismuth oxyiodide/fluorine-doped tin oxide(BiOI/FTO) electrodes using the cyclic voltammetry method
with the expectation of combining the excellent properties
of the polyacid(high electrocatalytic activity
continuous and reversible multi-electron transfer ability
and strong redox ability)for the matching between the polyacid energy level structure and the BiOI energy levels
in order to improve the performance of the BiOI/FTO electrodes and to enhance the photoelectric catalytic activity of the composite electrode.The detection of tetracycline hydrochloride is studied using H
2
V
12
O
31
/BiOI/FTO as the working electrode
and the intensity of the photoelectric response decreases with the increase of the concentration of tetracycline hydrochloride within the range of 0.1~1000μmol/L
which shows a good linear relationship
and the linear equation is
I
(μA cm
-2
)=2.843log
C
TC
(μmol/L)-10.417μA
the sensitivity is 113.153μA(mmol/L)
-1
cm
-2
and the limit of detection(LOD)(
S
/
N
=3)is 0.470μmol/L.The experimental results show that the sensor has a wide detection range for tetracycline and has good selectivity and stability.